KF, ISO-K, ISO-F and CF: which flange system do you need
A working comparison of the four common vacuum flange systems, their standards, size ranges and the adapters required to join them.
Flange standards are not alternative brands of the same part. They are different sealing systems, and joining two of them always costs an adapter and usually limits the whole assembly to the weaker end.
The four systems
| System | Sealing | Fastening | Standard | Typical DN |
|---|---|---|---|---|
| KF / ISO-KF | Elastomer O-ring on a centering ring | Wing-nut clamp | ISO 2861:2020 / DIN 28403 | 10–50 |
| ISO-K | Elastomer O-ring on a centering ring | Claw clamps around the flange | ISO 1609 / DIN 28404 | 63–1000 |
| ISO-F | Elastomer O-ring, bolted | Bolt circle through the flange | ISO 1609 / DIN 28404 | 63–1000 |
| CF / ConFlat | OFHC copper gasket against a knife edge | Bolt circle | ISO 3669:2017 | 10–250+ |
The three rules that decide most questions
1. KF and ISO have no overlapping DN. KF ends at DN50; ISO starts at DN63. There is no KF63-to-ISO63 straight connection. A reducer or a dedicated adapter is mandatory, and its conductance becomes the bottleneck.
2. ISO-K and ISO-F share one dimension series. At the same DN the flange faces are dimensionally compatible; only the fastening differs. A coupling flange with claws on one side and a bolt circle on the other joins them. The DN must still match.
3. Mixing CF with an elastomer-sealed system limits the whole assembly. CF uses a metal knife edge and a copper gasket, baked to temperatures no elastomer tolerates. A CF-to-KF adapter exists, but the system's achievable vacuum and bake-out temperature are then set by the elastomer end, not the CF end.
KF dimensions worth memorising
| DN | Flange OD A (mm) | Bore B (mm) | ISO preferred |
|---|---|---|---|
| 10 | 30.0 | 12.2 | Yes |
| 16 | 30.0 | 17.2 | Yes |
| 20 | 40.0 | 17.0 | No |
| 25 | 40.0 | 26.2 | Yes |
| 32 | 55.0 | 34.0 | No |
| 40 | 55.0 | 41.2 | Yes |
| 50 | 75.0 | 52.3 | Yes |
DN20 and DN32 are non-preferred sizes. Dimensions in published catalogues disagree for these two, so verify against the supplier drawing before machining a mating part.
Dimensions above are from public product catalogues citing ISO 2861:2020 / DIN 28403. They are reference values, not a substitute for the standard text or the supplier drawing.
Seal material sets the temperature ceiling
| Material | Typical service range | Reusable |
|---|---|---|
| NBR / Buna-N | -20 to 100 °C | Yes |
| FKM / Viton | -20 to 200 °C | Yes |
| FFKM / Kalrez | Ask the supplier | Yes |
| OFHC copper (CF) | -200 to 450 °C | No |
Published FKM bake-out limits vary between 150 °C and 200 °C. Use the conservative figure and confirm with the supplier datasheet. A copper gasket is single-use: a reused one is a leak.
Choosing by application
- Small pumps, lab lines, instruments: KF. Cheap, fast to assemble, adequate for rough and medium vacuum.
- Large chambers, forelines above DN63: ISO-K for frequent disassembly, ISO-F when the joint is permanent or vibration is present.
- UHV, bake-out, surface analysis: CF throughout. Do not introduce an elastomer seal into the baked section.
- A pump with a threaded port: none of the above applies until a thread-to-KF adapter is fitted. See the inlet fitting guide.
Note the distinction between leak rate and ultimate pressure: a typical elastomer-sealed KF joint is quoted around1 × 10⁻⁹ mbar·L/s, while catalogues list vacuum ranges down to roughly10⁻⁸ Torrfor elastomer seals and10⁻¹² Torrfor metal gaskets. These are different measurements and are not interchangeable.
Model and inlet data compiled from LeyNow official product pages (leynow.com) and public flange standards. Unverified fields are marked as per factory datasheet.